Safety electronic systems defined by standard DIN EN 60601-1 and EU guideline 93/42/EWG are used for medical applications. The REOMED series is a medical power supply certified by IEC60601-1:2005 (3rd edition), with a capacity range within 2200VA. REOMED devices can be customized and processed according to customer requirements. Special requirements include: additional shielding between primary and secondary windings, inrush current limiters (NTC or other electronic devices) and equipotential grounding pins according to IEC42801.

Compared with traditional transformers, the low loss and high efficiency performance of REOMED medical isolation transformers can meet up to 9 power supply circuits. REOMED Medical Power Supply's product portfolio offers RFI filters, surge protection and half-wave loss protectors with different parameters.

REOMED isolation transformers can be combined with REOMED ISOMONITOR insulation monitors to form a power supply and monitoring solution. The following monitor parameters and transformer parameters are listed separately and should correspond to specific components when selecting.

REOMED ISOMONITOR Monitor Performance Parameters
Power supply voltage230 V ~ 50/60 Hz +/- 10%
Power consumption1.3 VA
operating range230 V +/- 10%
Input/output voltage207 - 253 VAC
Trigger alarm value≤ 50 kΩ
response time< 2 Sec.
Alarm display signalLED green: Operation
LED yellow: insulation fault
LED yellow: Transformer temperature limit
acoustic signalA prominent and pulsing beep indicates an insulation fault; a continuous beep indicates an over-temperature alarm.
ambient operating temperature0 – +40°C
ambient relative humidity30 - 75 %
protection levelII
Entry protection levelIP 40
Dimensions [height x width x depth]125 x 67 x 40 mm
Connecting line 230V ACH 05 VV-F 3 G1, 0 with plug conforming to IEC 320 Cable length (standard) 0.75 m; optionally 2.5m
Cable/plug temperature monitoringBinder Series 682 angle plug, 2-pin 250 V AC/7A, cable length (standard) 0.75 m; optionally 2.5m
Related standardsEN60601-1, EN60601-1-1, EN60601-1-2, EN14971
Order number: REOMED ISOMONITOR with optional temperature monitoringATK129196910 ATK129196911

REOMED Isolation Transformers: Performance and Mechanical Parameters

Performance and mechanical structural parameters
Input voltage VOutput voltage VNominal power consumption VAIP rating
101 - 130 115 - 230 200 - 250100 - 130300~2200IP20
115 -230
200 - 250
Matching accessories
Technical illustration
Select image to view full size

REOMED Isolation Transformers: Dimensions and Weight Range

WidthheightDepthweight
155-235mm85-125mm240-325mm4.5 ~19kg